12
Reporting Temperature
App
lication Note 42
Output and Report Temperature in ITS-90
Calibration Reports show both sets of coeff.
Use IPTS-68 to compute salinity
SBE software does this automatically for you
What is difference?
IPTS-68 Versus ITS-90…1968 standards vs. 1990
IT
S-90 (1990) standards include water triple-point and gallium melt cell,
SPRT, and ASL F18 Temperature Bridge
Sea-Bird software and instrument converts between IPTS-68 and ITS-
90 according to the linear relationship:
T
68
= 1.00024 * T
90